Karl Fischer titrator selection guide

  Titrando® Titrinos® Titrino plus® Coulometers
  841 795 870 756 831
Modes of operation
Volumetric Karl Fischer (high moisture levels) Yes Yes Yes
Coulometric Karl Fischer (low moisture levels) Yes Yes
Karl Fischer detection limits 10 ppm 10 ppm 10 ppm 10 µg 10 µg
Karl Fischer back titration (with optional second buret) Yes
SET titration to a preset endpoint Yes
Measurement (pH, U/mV, T/°C) Yes
pH calibration with automatic buffer recognition Yes
Dosing elements (applies to volumetric KF only)
Built-in holder for two Dosinos® Yes
Integrated Dosimat for Exchange Unit Yes Yes (intelligent)
Maximum number of titrating burets 4 1 1
Intelligent Dosing Unit with built-in SmartChip™ Yes Yes
Stores titrant standardization history Yes Yes
Operation
Color, LCD Touch Control (9 cm x 12 cm) Yes
8-line, B/W LCD with separate keypad Yes Yes Yes Yes
Real-time curve display Yes Yes Yes Yes Yes
Storage for titration results including curve Yes No
Internal storage for methods Yes Yes Pre-programmed methods only Yes Yes
Memory card storage for results, methods Yes Methods only
Traceability and security
Reagent monitoring and traceability (intelligent dosing devices) Yes Yes
User name Yes Yes Yes Yes
Login with password protection Yes
Electronic signature Yes
Audit trail Yes
Stirrer speed control in method Yes Yes
Automatic stirrer on/off Yes Yes Yes Yes Yes
FDA 21 CFR Part 11 Compliant Yes With software only With software only With software only
Monitoring
Confirms correct reagent is used Yes Yes
Confirms correct hardware is connected Yes Yes
Checks sample size limits Yes Yes Yes Yes Yes
Checks result limits Yes Yes Yes Yes Yes
Titrant standardization intervals Yes Yes
Validation/service intervals Yes Yes Yes Yes Yes
Electrode calibration intervals Yes
Connections
MSB port for Dosimats, Dosinos, stirrers 4 1 (stirrer only)
USB port (sample changer, balance, printer, barcode reader) 2 1 (expandable)
RS-232 (balances, printer, barcode reader) 2 (optional) 2 2 (optional) 2 2
Remote port Optional Yes Yes Yes Yes
Memory card for methods, results, keycard (user identification) Yes, PCMCIA Methods only USB methods/config only
Reagent exchange with Dosino Yes No Yes Yes
Built-in thermal printer No Yes  
Connects to Karl Fischer ovens Yes Yes Yes Yes Yes
Connects to homogenizer system Yes Yes Yes
Connects to Tiamo™ titration software Yes Yes No Yes Yes