| Heating: |
Two individually heated aluminum heating blocks |
| Max. number of samples: |
8 (4 in each block) |
| Temperature range: |
50°C to 220°C, can be set in steps of 1°C |
| Temperature correction: |
–9.9°C to +9.9°C, can be set in steps of 0.1°C |
| Max. deviation of set temperature: |
<0.3°C |
| Reproducibility of set temperature: |
<0.2°C |
| Stability of temperature (max. temperature fluctuation): |
<0.1°C |
| Temperature difference between different measuring positions: |
<0.3°C |
| Heating time from 20°C:
to 120°C
to 220°C |
Approx. 45 min
Approx. 60 min |
| Gas supply: |
External supply |
| Air flow range: |
4 L/h to 15 L/h |
| Max. deviation of flow: |
±10% |
| Conductivity measurement: |
0 µS/cm to 500 µS/cm |
| Interface: |
RS-232C |
| Software: |
Metrodata 763 PVC Thermomat software 1.0 |
| Dimensions (W x H x D): |
405 x 353 (with accessories) x 402 mm
15 x 14 (with accessories) x 16 in |
| Weight: |
26.2 kg/57.6 lb |