Technical Specifications

Catalog No. 027440000 027440300 027440400 027440100
Model XT 150 XT 250 W XT 350 HW XT 750
Operating temperature range (°C):

–45 to 200

–45 to 200

–45 to 200

–45 to 200

Ambient termperature (°C):

5 to 40 5 to 40 5 to 40 5 to 40
Resolution of setting (°C): 0.01 0.01 0.01 0.01
Cooling: Air Water Water Air
Filling volume, min. (L): 2.6 2.6 5.3 5
Filling volume of expansion vessel (L): 5.5 5.5 6.7 6.7
Refrigerant: R404A R404A R404A R404A
Water-cooling requirements, suggested at 15 °C, 3 bar (L/H): 300 540
Water-cooling connector fittings: G 3/4" A G 3/4" A

Cooling capacity at 20 °C ambient temperature (pump level setting at 4 or 2), heat transfer oil (kW):
  300°C:
  250°C:
  205°C:
  200°C:
  150°C:
  100°C:
  50°C:
  20°C:

P2






1.5b
1.5b
1.5b
1.5b
1.5b

P2






1.9b
2.1b
2.1b
2.1b
2.1b

P4



12.0
12.0
12.0
12.0a
9.0a
6.0a
3.3a
3.1a

P4






7.0b
7.0b
7.0b
7.0b
7.0b

ethanol (kW)
  Ref. 20°C:
  10°C:
  0°C:
  –10°C:
  –20°C:
  –30°C:
  –40°C:
  –45°C:
  –50°C:


1.5
1.3
1.1
1.0
0.62
0.28
0.06


2.1
1.8
1,3
1.0
0.62
0.28
0.06


3.1
2.9
2.5
1.3
1.2
0.7
0.25
0.15
0.02

6.7
6.1
4.8
3.4
2.2
1.25
0.6

0.3
Temperature stability (DIN 12876) at –10°C (±°C): 0.05 0.05

0.1

0.05
Measurement taken with ethanol, with given external volume (L): 2 2 5 5
Heating capacity, max. (kW): 3.5 3.5 3.5 5.2
Pumping specifications taken with water at 20°C
Pump flow, max. (L/min): 45 45 45 45
Pump pressure, max. (bar): 2.9 2.9 2.9 29
Pump connections: M30 x 1.5 (DN 20) M30 x 1.5 (DN 20) M30 x 1.5 (DN 20) M30 x 1.5 (DN 20)
Total dimensions, W x D x H (in/cm): 13.2 x 21.6 x 26.0/
33.5 x 55.0 x 66.0
13.2 x 21.6 x 26.0/
33.5 x 55.0 x 66.0
18.1 x 21.6 x 50.6/
46.0 x 55.0 x 128.5
18.1 x 21.6 x 50.6/
46.0 x 55.0 x 128.5
Power consumption (kW): 3.68 3.68 3.68 7.8
Weight (lb/kg): 191/87 198/90 330/150 341/155
Electrical requirements (V/Hz): 220/60 220/60 220/60 208–220/60, 3-phase
         
Catalog No. 027440500 027440200 027440600  
Model XT 950 W XT 750 H XT 1850 W  
Operating temperature range (°C):

–50 to 200

–50 to 200

–50 to 200

 

Ambient termperature (°C):

5 to 40 5 to 40 5 to 40  
Resolution of setting (°C): 0.01 0.01 0.01  
Cooling: Water Air Water  
Filling volume, min. (L): 5 5 9,0  
Filling volume of expansion vessel (L): 6.7 6.7 17.4  
Refrigerant: R404A R404A R404A  
Water-cooling requirements, suggested at 15 °C, 3 bar (L/H): 1,300 1,320  
Water-cooling connector fittings: G 3/4’’ A G 1’’ A  

Cooling capacity at 20 °C ambient temperature (pump level setting at 4 or 2), heat transfer oil (kW):
  300°C:
  250°C:
  205°C:
  200°C:
  150°C:
  100°C:
  50°C:
  20°C:

P4






7.3b
7.3b
9.0b
9.0b
9.0b

P4



5.5
4.5
3.5
7.0a
7.0a
7.0a
7.0a
7.0a

P4






18.5a
18.5a
18.5a
18.3a
18.5a

 

ethanol (kW)
  Ref. 20°C:
  10°C:
  0°C:
  –10°C:
  –20°C:
  –30°C:
  –40°C:
  –45°C:
  –50°C:


9.0
7.1
6.7
4.7
3.0
1.8
1.0

0.3


6.7
6.1
4.8
3.4
2.2
1.25
0.6

0.3


18.5
12.5
10.3
7.7
5.9
3.8
2.2

1.2
 
Temperature stability (DIN 12876) at –10°C (±°C): 0.1 0.1

0.3

 
Measurement taken with ethanol, with given external volume (L): 5 5 10  
Heating capacity, max. (kW): 5.2 5.2 10.4  
Pumping specifications taken with water at 20°C  
Pump flow, max. (L/min): 45 45 90  
Pump pressure, max. (bar): 2.9 2.9 5.8  
Pump connections: M30 x 1.5 (DN 20) M30 x 1.5 (DN 20) M38 x 1.5 (DN 25)  
Total dimensions, W x D x H (in/cm): 18.1 x 21.6 x 50.6/
46.0 x 55.0 x 128.5
18.1 x 21.6 x 50.6/
46.0 x 55.0 x 128.5
27.6 x 21.6 x 63.0/
70.0 x 55.0 x 160.0
 
Power consumption (kW): 7.8 7.8 15.4  
Weight (lb/kg): 352/160 352/160 550/250  
Electrical requirements (V/Hz): 208–220/60, 3-phase 208–220/60, 3-phase 208–220/60, 3-phase  
a Measurements taken with Ultra 350 bath liquid.