| Model |
WKL 230 |
WKL 600 |
WKL 900 |
WKL 1000 |
|
Catalog
No. |
027642608 |
027642659 |
027642888 |
027642896 |
| Working temperature range (°C): |
–10 to
40 |
–25
to 40 |
–20 to
40 |
–10 to
40 |
| Temperature control (±°C): |
0.1 |
0.1 |
0.1/1 0 |
1.0 |
| Temperature stability (+°C): |
0.5 |
1.0 |
1.0 |
0.1 |
| Cooling output at 20°C (W): |
230 |
650 |
950 |
1100 |
| Pump pressure max. (bar): |
1.5 |
1.0 |
1.0 |
1.0 |
| Pump flow max. (L/min): |
8 |
30 |
30 |
30 |
| Filling volume (L): |
6 |
12 |
12 |
12 |
| Noise level (dB): |
47 |
53 |
54 |
50 |
Dimensions,
D x W x H (in):
D x W x H (cm): |
13.8 x 7.9 x 19.7
35.0 x 20.0 x 50.0 |
13.8
x 18.7 x 23.0
35.0 x 48.0 x 58.4 |
13.8 x 18.7 x 23.0
35.0 x 48.0 x 58.4 |
13.8 x 18.7 x 23.6
35.0 x 48.0 x 60.0 |
| Voltage requirements: |
115 V, 60 Hz |
230
V, 60 Hz |
208 V–230
V, 60 Hz |
115 V, 60 Hz |
| |
|
|
|
|
| Model |
WKL 1200 |
WKL 2200 |
WKL 2200
to –25°C |
WKL 3200 |
|
Catalog No. |
027642900 |
027643159 |
027643167 |
027643205 |
| Working temperature range (°C): |
–10 to
40 |
–10 to
40 |
– 25 to
40 |
–10 to
40 |
| Temperature control (±°C): |
0.1 |
0.1 |
0.1 |
0.1 |
| Temperature stability (+°C): |
0.5 |
1.0 |
1.0 |
1.0 |
| Cooling output at 20°C (W): |
1,200 |
2,200 |
2,200 |
3,500 |
| Pump pressure max. (bar): |
3.2 |
3.2 |
3.2 |
3.2 |
| Pump flow max. (L/min): |
40 |
40 |
40 |
40 |
| Filling volume (L): |
23 |
23 |
23 |
45 |
| Noise level (dB): |
60 |
59 |
59 |
60 |
Dimensions,
D x W x H (in):
D x W x H (cm): |
17.7 x 21.7 x 31.1
55.0 x 45.0 x 79.0 |
17.7 x 21.7 x 31.1
55.0 x 45.0 x 79.0 |
17.7 x 21.7 x 31.1
55.0 x 45.0 x 79.0 |
25.6 x 21.7 x 37.4
65.0 x 55.0 x 95.0 |
| Voltage requirements: |
208 V–230
V, 60 Hz |
208 V–230
V, 60 Hz |
230 V, 60 Hz |
230 V, 60 Hz |
| |
|
|
|
|
| Model |
WKL 3200 to
–25°C |
WKL 4600 |
WKL 4600 to
–25°C |
WKL 7000 |
|
Catalog No. |
027643213 |
027643256 |
027643264 |
027643337 |
| Working temperature range (°C): |
– 25 to
40 |
–10 to
40 |
–25 to
40 |
–30 to
40 |
| Temperature control (±°C): |
0.1 |
0.1 |
0.1 |
0.1 |
| Temperature stability (+°C): |
1.0 |
1.0 |
1.0 |
0.5 |
| Cooling output at 20°C (W): |
3,500 |
4,600 |
4,600 |
7,000 |
| Pump pressure max. (bar): |
3.2 |
3.2 |
3.2 |
6.0 |
| Pump flow max. (L/min): |
40 |
40 |
40 |
60 |
| Filling volume (L): |
45 |
45 |
45 |
45 |
| Noise level (dB): |
60 |
63 |
61 |
65 |
Dimensions,
D x W x H (in):
D x W x H (cm): |
25.6 x 21.7 x 37.4
65.0 x 55.0 x 95.0 |
25.6 x 21.7 x 37.4
65.0 x 55.0 x 95.0 |
25.6 x 21.7 x 37.4
65.0 x 55.0 x 95.0 |
33.5 x 26.1 x 37.8
85.0 x 67.0 x 97.0
|
| Voltage requirements: |
208 V, 60 Hz |
208 V, 60 Hz |
208 V, 60 Hz |
460 V, 60 Hz |
| |
|
|
|
|
| Model |
WKL 10000 |
|
|
|
|
Catalog No. |
027643370 |
|
|
|
| Working temperature range (°C): |
–30 to
40 |
|
|
|
| Temperature control (±°C): |
0.1 |
|
|
|
| Temperature stability (+°C): |
0.5 |
|
|
|
| Cooling output at 20°C (W): |
10,000 |
|
|
|
| Pump pressure max. (bar): |
6.0 |
|
|
|
| Pump flow max. (L/min): |
60 |
|
|
|
| Filling volume (L): |
45 |
|
|
|
| Noise level (dB): |
69 |
|
|
|
Dimensions,
D x W x H (in):
D x W x H (cm): |
41.0 x 30.0 x 43.7
105.0 x 77.0 x 112.0 |
|
|
|
| Voltage requirements: |
440 V–480 V, 60 Hz
|
|
|
|